Reference
A pH sensor, measuring hydrogen ion concentration, indicates acidity or alkalinity. It comprises a glass and reference electrode. For initial use, soak the probe in 3NKCL solution for 8 hours. Maintain the glass bubble's moisture, avoid hard objects, and after use, store it with 3mol/L KCL solution. Regular cleaning with specific solutions is crucial to maintain accuracy and prevent contamination. If sensitivity drops, clean the electrode with 4%HF and 0.1MOL/L HCL. Avoid using soluble polycarbonate cleaning liquids as they can damage the probe. Follow these steps to ensure accurate and reliable pH measurements. It's like keeping a delicate instrument in top condition.
Principle
A pH sensor measures the hydrogen ion concentration in water-based solutions, indicating its acidity or alkalinity expressed as pH. It usually has a glass electrode and a reference electrode.
Other Supplementary Information
Put the probe of the sensor into the solution to be tested. Connect all parts together as shown below, power up and then the readings will be displayed on the OLED module.


Precautions:
- For the first time use or re-use after a long time, please immerse the sensor probe in 3NKCL solution for 8 hours.
- The glass bubble in the head of the sensor shall not come in contact with any hard objects since any breakage or rubbing will invalidate the probe.
- After you finished the measurement, please put a small amount of 3mol/L KCL solution inside the protective cap and put the cap on the sensor. The glass bubble should always be kept moist.
- The plug of the pH sensor must be kept clean and dry, and the output ends shoud never be short-circuit so as to prevent the inaccurate measurement and product failure.
- Do not immerse the probe in distilled water, protein, acid fluoride solution for a long time, and keep the sensor away from silicone oil.
- After long-term use, if you found that the percentage of the theoretical slop of the product reduced, you can soak the bottom of the electrode in 4%HF(hydrofluoric acid) for 3-5 seconds, wash with distilled water, and then soak it in the 0.1MOL/L HCL solution for a few hours, rinse clean with deionized water.
- The presence of contaminants in the tested solution may passivate the sensitive bulb of the probe, as a result, the sensitivity and accuracy of the sensor will be reduced. In this case, we suggest that you clean the product with an appropriate solution according to the nature of the contaminants.
- The soluble polycarbonate cleaning liquid, such as carbon tetrachloride, trichloroethylene, four furans and so on, may stain the glass bubble of the probe and then damage the electrode, please use them with great caution!