This article delves into the communication protocols for the SCA100T-D2 chip, detailing commands for outputting values, the API, and its application in precision leveling instruments.

Command A(Output the processed value)

Command:(Hexadecimal)

Header Command Check sum
55 AA 01 FE

Return:(hexadecimal)

55 AA FF 01 19 00 00 2D 21 14
Hearer X sign bit positive 00,negative FF First interger Second decimals Y sign bit positive 00,negative FF First interger Second decimals Temperature Check sum

Command B(Output original value )

Command:(Hexadecimal)

Header Command Check sum
55 AA 02 FD

Return:(Hexadecimal)

55 AA 04 02 04 09 A4 50
Header X high 8 bits X low 8 bits Y high 8 bits Y low 8 bits Temperature Check sum

Principle

This gradienter module is based on the SCA100T-D2 chip, which provides instrumentation grade performance for leveling applications.The measuring axis of the sensing elements are parallel to the mounting plane and orthogonal to each other. Benefiting from its low temperature dependency, high resolution and low noise, together a with robust sensing element design, the SCA100T the ideal choice for leveling instruments.