Gravity: Analog Dissolved Oxygen Sensor

SKU: SEN0237-A

Gravity: Analog Dissolved Oxygen Sensor is an open-source, Arduino-compatible kit designed to measure dissolved oxygen levels in water. Featuring a galvanic probe requiring no polarization time, replaceable filling solution and membrane cap for low maintenance, and easy plug-and-play integration, this product is ideal for aquaculture, environmental monitoring, and scientific research.

Front of Arduino dissolved oxygen sensor module for water quality monitoring-DFRobot product sku:SEN0237-A image.
  • Docs
  • Tech Specs
  • Projects
  • 1. Getting Started

    This article explains how to properly prepare a dissolved oxygen probe by filling it with 0.5 mol/L NaOH solution, covering the steps to avoid errors such as bubble formation and ensuring accurate measurements.

    2. Example Code for Arduino-Probe Calibration

    This article offers example code and comprehensive instructions for calibrating probes to accurately measure dissolved oxygen values. It explains single-point and two-point calibration methods, providing guidance on hardware and software setup. Readers will learn how to achieve precise sensor readings under stable and varying temperature conditions.

    3. Example Code for Arduino-Dissolved Oxygen Measurement

    This tutorial will show how to use this dissolved oxygen sensor kit. The dissolved oxygen probe is a precision electrochemical sensor. Please pay attention to the usage details.

    Reference

    The article explores the calibration principles necessary for accurate dissolved oxygen measurement, focusing on the relationship between temperature and saturation voltage, and the need for temperature compensation curves to ensure measurement precision.

    Specification

    Dissolved Oxygen Probe

    Parameter Details
    Type Galvanic Probe
    Detection Range 0~20 mg/L
    Temperature Range 0~40 ℃
    Response Time Up to 98% full response, within 90 seconds (25℃)
    Pressure Range 0~50 PSI
    Electrode Service Life 1 year (normal use)
    Membrane Cap Replacement Period: 1~2 months (in muddy water); 4~5 months (in clean water)
    Filling Solution Replacement Period: Once every month
    Cable Length 2 meters
    Probe Connector BNC

    Signal Conditioner Board

    Parameter Details
    Supply Voltage 3.3~5.5V
    Output Signal 0~3.0V
    Cable Connector BNC
    Signal Connector Gravity Analog Interface (PH2.0-3P)
    Dimension 42mm /* 32mm / 1.65 /* 1.26 inches

    Pinout

    Dissolved_Oxygen_Sensor_Pinout

    No. Label Description
    1 A Analog Signal Output (0~3.0V)
    2 + VCC (3.3~5.5V)
    3 - GND
    4 BNC Probe Cable Connector

    Was this article helpful?

    TOP